Minimizing matrix effects while preserving throughput in LC–MS/MS bioanalysis


Background: Phospholipids are known to cause matrix effects in LC–MS analysis and are not effectively removed by one of the most common method of sample preparation: organic solvent protein precipitation. The objective of this research is to minimize phospholipid interferences chromatographically. Results: In this article we examine several chromatographic approaches and highlight the method we developed that allows for the rapid gradient separation of model drug molecules from phospholipids. Conclusion: The new approach (which utilizes a mixture of methanol and acetonitrile as the organic mobile phase on a 2.1 × 20 mm C18 column) minimized phospholipids-related matrix effects in the...

To view this content, please register now for access

It's completely free